Calibrate electronic process instruments such
as controllers, transmitters, transducers or recorders with Model
CAL10 Thermocouple Calibrator. Model CAL10 is designed to independently
measure and simulate dc mV or V and common thermocouple signals
(Type J, K, T, E, J (DIN), and T (DIN) or measure current.
Programmable functions, such as square root extraction for flow
signals, ramp/step increment value, selectable engineering units,
isolated signal conversion or automatic changing of the output
value simplify many calibration tasks. Calibration values can
be stored and recalled using the Low, Mid, or High keys. Cold
junction compensation can be selected for automatic or manual.
Program the automatic shut-off to conserve battery life.
The Model CAL10 calibrator features 500 VDC input/output isolation
and reverse polarity protection. The unit includes three 9V batteries,
two sets of dual test leads, a dual binding post adapter, carrying
case, and instruction manual.
|
Input Ranges: Voltage: -10 to 24 mV DC/-10 to 120 mV
DC (autoranging), 0 to 12 VDC; Current: 0 to 24 mA DC; Temperature:
see table below.
Accuracy: Voltage and Current: ±0.025% of range; Temperature:
see table below.
Output Ranges: 0 to 120 mV DC (2K Ω min. load); 0 to 12 VDC (2K Ω min.
load); Thermocouple Types J, K, T, E, J (DIN), and T (DIN) similiar to input
ranges.
Display: 24-character LCD, 3/16" (5 mm) character height.
Indication: Up to 5 digits depending on range.
Resolution: 0.1°C/F.
Response Time: 5 seconds.
Ambient Operating Temperature: -4 to 113°F (-20 to 45°C).
Electrical Connections: Standard banana jacks on 3/4" (19 mm) centers.
Power Requirements: Three 9V alkaline (included).
Battery Life: 100 hours, approximate.
Weight: 1.15 lbs. (0.52 kg).
Agency Approvals: CE.
|